CALibration[<chan>]:STORe

:STORe

CALibration[<chan >]:STORe stores the current calibration constants into non-volatile calibration memory.

Parameters

Parameter

Parameter

Range of

Default

Name

Type

Values

Units

 

 

 

 

chan

numeric

1 or 2

none

 

 

 

 

Comments Executable when initiated: No

Coupled Command: No

CALibration:SECure:STATe must be OFF before executing this command.

Related commands: CALibration[<chan>]:DATA, CALibration:SECure:STATe, CALibration:COUNt?, CALibration:STORe:AUTO

*RST Condition: none

Example Sending and storing an array of new cal constants

ASSIGN @X TO 70905;FORMAT OFF

 

Turn format off for array data

OUTPUT 70905 USING "#,K";"CAL:DATA #3124"

 

Specify 124 bytes coming (62

 

constants)

OUTPUT @X;Array(*),CHR$(10),END

Send the array of calibration constants

OUTPUT 70905;"CAL:SEC:OFF,E1429"

Disable security

 

OUTPUT 70905;"CAL:STOR"

Store the calibration data in non-volatile

 

RAM