HP E1429A manual ∙ Related Commands ARMSTARtSLOPen , ARMSTARtLEVelchan

Models: E1429A

1 420
Download 420 pages 44.04 Kb
Page 197
Image 197

ARM[:STARt]:SOURce[<n>]

Parameters

Parameter

Parameter

Range of

Default

Name

Type

Values

Units

 

 

 

 

n

decimal

12

none

 

 

 

 

source

discrete

BUSECLTrg0

none

 

 

ECLTrg1IMMediate

 

 

 

EXTernal1INTernal[1]

 

 

 

INTernal2HOLD

 

 

 

TTLTrg0 to TTLTrg7

 

 

 

 

 

Choice IMMediate is only available for ARM:STARt:SOURce1.

Comments Executable when initiated: No

Coupled Command: Yes. The command is coupled to the ARM[:STARt]:SLOPe[<n >], ARM[:STARt]:LEVel:NEGative, and ARM[:STARt]:LEVel:POSitive commands.

The active edge for the EXTernal1 source is specified by the ARM:STARt:SLOPe[<n >] command.

The active edge(s) for level detection (ARM:STARt:SOUR:LEVel<chan >) is specified by the ARM:STARt:SLOPe[<n >] command.

Related Commands: ARM:STARt:SLOPe[<n >], ARM:STARt:LEVel[<chan >]

*RST Condition: ARM:STARt:SOURce1 IMMediate, ARM:STARt:SOURce2

HOLD

Example Setting two arm start sources

ARM:SOUR1 EXT1

Set one arming source to be the front panel

 

"Ext 1" BNC.

ARM:SOUR2 ECLT0

Set the second arming source to be the

 

ECLTRG line.

ARM Subsystem

Command Reference 197

Page 197
Image 197