Processor Humidity Specifications

Specifications

Processor Humidity Specifications

If the humidity is low (approximately 15 - 20%), check the chemistry levels in the tank every 2 hours. Make sure that the chemistry in each processor tank is at the top of the overflow pipe. If necessary, add water to reach the correct level.

overflow pipe

IMPORTANT: Operate the processor’s main drive for approximately 15 to 30 minutes after topping off the tanks.

Wash Chemicals

 

Maximum Environmental Temperature

Wash Chemical

 

 

 

 

 

 

 

 

 

 

 

 

 

 

27

o

Celsius (80

o

Fahrenheit)

Stabilizer Wash

 

 

 

 

 

 

27o

30oCelsius (80o— 86o Fahrenheit)

Water Wash

Telephone line

A telephone line (with a RJ11C modular phone jack) must be installed within

3 m (10 ft.) of the printer to support the remote diagnostics. We recommend a high-grade analog service line.

Densitometer

An X-RITE DTP— 36 Densitometer is required to calibrate the printer and for process control. It is available through Kodak. See “Accessories” on page A-1.