Kodak 20P Calibration Error Codes, Diagnostics and Troubleshooting, Error Code /Status, Message

Models: 20P

1 198
Download 198 pages 51.29 Kb
Page 110
Image 110
Calibration Error Codes

Diagnostics and Troubleshooting

Calibration Error Codes

Error

Error Code /Status

Possible Cause/Subsystem

 

Possible Solution

Code

Message

 

 

 

 

 

 

 

 

1

The following file is

The system is trying to access a

Close any other applications that

 

locked:_filename_.

file that is currently in use.

may be accessing the file.

 

 

 

 

2

The following file is

The system is trying to access a

Change the protections on the file

 

protected:_filename_.

file that is protected.

to grant access to the user.

 

 

 

 

3

The following file was not

The system was trying to access

Verify that the filename on one of

 

found:_filename_.

a file that it could not find.

the configuration dialog windows is

 

 

 

correct.

 

 

 

 

 

4

Error:_oserror_occurred

An operating system error

Rerun the application.

 

during operation:_operation_

occurred.

• Reboot the operating system

 

on file:_filename_.

 

 

 

 

 

 

 

5

The following file is in an

The file does not conform to the

Verify the file format.

 

incorrect format:_filename_.

application specification.

 

 

 

 

 

 

 

6

The following file already

The system is trying to save to a

Delete the existing file.

 

exists:_filename_.

file that already exists.

Change the filename.

 

 

 

 

7

There is not enough space to

The system is trying to save a file

Remove any unnecessary files to

 

create:_filename_.

but there is not enough disk

increase free disk space.

 

 

space.

 

 

 

 

 

 

8

There are too many files to

Too many files open.

Close some files and or

 

open _filename_.

 

applications.

 

 

 

 

9

You do not have permission

The file permission is invalid.

Rerun the application.

 

to open file _filename_.

 

 

 

 

 

 

 

10

Error copying print LUT to

Folder doesn’t exist or is read-

Modify protection on folder.

 

DP2 folder (can’t open

only.

 

 

 

_filename_.

 

 

 

 

 

 

 

11

Error creating DP2 LUT

A folder of the same name may

Change the selected folder name.

 

folder

exist.

 

 

 

 

 

 

101

No response from

The device is not responding to

Verify that the device is connected

 

device:_devicename_.

the system.

and on-line.

 

 

 

 

102

Device not

The system was trying to access

Verify that the device name on one

 

found:_devicename_.

a device that it could not find.

of the configuration dialog

 

 

 

windows is correct.

 

 

 

 

103

The following device is

The device is not ready.

Wait several minutes and try the

 

indicating it is not

 

activity again.

 

ready:_devicename_.

 

 

 

 

 

 

 

104

The following device is

The device is busy.

Wait several minutes and try the

 

busy:_devicename_.

 

activity again.

 

 

 

 

105

The following device is not a

The port name in the

Change the port name in the

 

TTY port:_devicename_.

configuration file does not refer to

configuration file to a valid serial

 

 

a valid serial port.

port.

 

 

 

 

 