Kodak 20P manual Editing Information on the Density Source Tab, Additional Calibration Information

Models: 20P

1 198
Download 198 pages 51.29 Kb
Page 158
Image 158
Editing Information on the Density Source Tab

Additional Calibration Information

Editing Information on the Density Source Tab

The Density Source tab specifies how the calibration application is to receive density data. The values displayed in the example are the default settings.

Density Source

Description

Attribute

 

 

 

 

 

Density Source

Indicates the source (densitometer or file) of the density data.

 

 

Density Source Port

Indicates the name of the RS-232 port that the densitometer is

 

connected to. This port is the source of the density data. The name of the

 

port is specific to the computer system platform.

 

 

Density Source File

Indicates the filename for the source of density data. This is only used if

 

the Density Source is a file. If you designate a file as the density source,

 

this is the file name you use for “Creating a Density File for Use with

 

Calibration” on page C-32.

 

 

To edit the configuration information on the Density Source tab:

1.Use the drop-down lists to change the Density Source information as needed.

2.If you have completed all editing of configuration information (including information on the other tabbed screens), click OK to save the changes and close the Edit Configuration screen.

NOTE:Click Cancel to cancel all changes made to the Edit Configuration screen and close the Edit Configuration screen.

3.To continue making changes to the Edit Configuration, click on another tab and make changes as needed.

C-24

September 2000

Page 158
Image 158
Kodak 20P manual Editing Information on the Density Source Tab, Additional Calibration Information, Description, Attribute