210OUTPUT OStimulus;"*IDN?"
220ENTER OStimulus;Id$
230PRINT Id$
240PRINT
250!
260PRINT "Counter Used . . . II
270OUTPUT OResponse;"*IDN?"
280ENTER QResponse;Id$
290PRINT Id$
300PRINT
310!
320OUTPUT OStimulus;":OUTPUT ON"
330!
340PRINT
350PRINT "INPUT [mvI","OUTPUT [kHz]"
360PRINT II
370PRINT
380!
390FOR Testpoint=First TO Last
400OUTPUT OStimulus;":SOURCE:VOLT ";VAL$(Testpoint/lOOO);";*OPC?"
410ENTER OStimulus;Dummy
420OUTPUT QResponse;":MEAS:FREQ?"
430ENTER OResponse;Reading
440PRINT Testpoint,Reading/lOOO
450NEXT Testpoint
460!
470OUTPUT QSource;":OUTPUT OFF"
480END
Program Comments. Lines 20 to 70: Declare variables and I/O paths for instruments. I/O paths let you use a name for an instrument in OUTPUT and ENTER statements,instead of a numeric address.
80 to 100: Assign values to the input test limits in mV.
110 to 130: Clear the instrument
140 to 160: Reset each instrument to a known measurement state.
170 to 190: Print the test report title.
200 to 310: Query measurement instruments’ identifications for test traceability.
320 to 330: Connect the source output signal to the output terminals.
340 to 380: Print results table header.
390 to 460: This is the main measurement loop. Line 400 contains two commands. :SOURce:VOLT sets the output level of the source. *OPC? is used to signal that the preceding command has finished executing. To make an accurate measurement,the source output must be allowed to settle. When the output has settled, *OPC? places
Getting Started Programming