H

HP-IB Address

HP-IB Menu

To set the synthesizer’s HP-IB address, refer to “Address” in this manual.

Function Group

Menu Map

Description

SYSTEM

8

This softkey reveals the softkeys in the HP-IB control menu.

Adrs Menu

 

 

Reveals

the softkeys that

 

 

 

allow HP-IB addresses to be

 

 

 

changed.

 

Programming

Language

Analyzr

Sets analyzer as the external

 

 

 

interface

language.

Progranrming

Language

CIIL

Sets CIIL as the external

 

 

 

interface

language.

Programming

Language

SCPI

Sets SCPI as the external

 

 

 

interface

language.

Three different programming languages are available:

nSCPI, Standard Commands for Programmable Instruments, is the instrument control programming language developed by Hewlett-Packard to conform to the IEEE 488.2 standard (replacing IEEE 728-1982). The IEEE 488.2 standard provides codes, formats, protocols, and common commands that were unavailable in the previous standard.

nAnalyzer is the programming language compatible with the

HP 8340/41 synthesized sweepers system language and many network analyzers.

nCIIL, Control Interface Intermediate Language, is the instrument control programming language used in option 700 synthesizers. Option 700 synthesizers are M.A.T.E. (Modular Automatic Test Equipment) compatible.

HP 8360

Operating and Programming Reference H-l