Chassis Kits

Rack Mount Slide Kit (Option 806)

Option 806 synthesizers are supplied with rack mount slides and the necessary hardware to install them on the synthesizer. The following table itemizes the parts in this kit.

Table 3-3. Rack Mount Slide Kit Contents

QuantityDescription

Rack Mount Kit (Includes the following parts)

2Rack Mount Flanges

8Screws

Slide Kit Jncludes the following parts)

2 Slide Assemblies

4Screws (Inner Slide Assembly)

8Screws (Outer Slide Assembly)

8Nuts (Outer Slide Assembly)

Slide Adapter Kit (NON-HP,includes the following parts)

4Adapter Brackets

4Adapter Bar

8Screws (Bracket to Bar)

8Nuts (Bracket to Slide Assembly)

CAUTION

Ventilation Requirements: When installing the instrument in a

 

cabinet, the convection into and out of the instrument must not be

 

restricted. The ambient temperature (outside the cabinet) must be

 

less than the maximum operating temperature of the instrument

 

by 4 “C for every 100 watts dissipated in the cabinet. If the total

 

power dissipated in the cabinet is greater than 800 watts, then forced

 

convection must be used.

 

 

Installation Procedure

1.Refer to Figure 3-3. Remove handle trim strips.

2.Remove four screws per side.

3.Using the screws provided, attach the rack mount flanges to the outside of the handles.

4.Remove the side straps and end caps.

5.Remove the bottom and back feet and the tilt stands.

3-10 INSTALLATION

Specifications