Start SweepTrigger Auto

Function Group

Menu Map

Description

Start Sweep Trigger Bus

SWEEP

7

When this softkey is selected, the synthesizer automatically triggers a sweep. This is the fastest way to accomplish the sweep-retrace cycle. An asterisk next to the key label indicates that this feature is active.

Programming Codes SCPI: TRIGger:SOURce IMMediate

Analyzer: Tl

See Also (CONT), (-1, Sweep Menu

Start Sweep

Trigger Bus

Function Group

Menu Map

Description

SWEEP

7

When this softkey is selected, the synthesizer waits for an HP-IB trigger to trigger a sweep. An asterisk next to the key label indicates that this feature is active.

Programming Codes

See Also

SCPI: TRIGger:SOURce BUS

Analyzer: TS

(CONT), (SINGLE), Sweep Menu

HP 8360

Operating and Programming Reference S-63