HP 8360
User’s Handbook
SCPI COMMAND SUMMARY
•SWEep:MANual:POINt <num>lMAXimumlMINimum
•SWEep:MANual:POINt? [MAXimumlMINimum]
Sets and queries the step point number to go to and lock. The value is a unitless value that is limited between 1 and the number of points requested. This command has no effect on the instrument unless: the sweep mode is set to manual and the sweep generation is set to stepped mode. *RST value is 1.
•SWEep : MANual [ : RELat ive] <mm>
•SWEep : MANual [ : RELat ive] ?
Sets and queries a percent of sweep to go to and lock. This command has no effect unless: the sweep mode is set to manual and the sweep generation is set to analog. *RST value is 0.50.
•SWEep:MARKer:STATe ONlOFFlllO
•SWEep:MARKer:STATe?
Sets and queries the state of marker sweep. When ON, the frequency sweep limits are taken to be the positions of marker 1 and marker 2. *RST value is 0.
•SWEep : MARKer : XFER
This transfers the values of marker 1 and marker 2 frequencies into start and stop frequency.
•SWEep:MODE AUTOlMANual
•SWEep : MODE?
Selects and queries the manual sweep mode switch.
A U T O The sweep is under the control of the INIT and SWEEP subsystems.
MANual FREQ:MANual, SWEep:MANual[:RELative], and SWEep:MANual:POINt control the output.
*RST value is AUTO.
•SWEep:POINts <num>lMAXimumlMINimum a SWEep : POINts? [MAXimum
I MINimum]
Sets and queries the number of points in a step sweep. When points is changed, SWEep : STEP is modified by the equation
STEP = SPAN/POINTS.
Span is normally an independent variable but is changed to STEP x POINTS if both of these parameters are changed in the same message. *RST value is 11.
•SWEep : STEP Cnum> [f req suffix] IMAXimum I MINimum
•SWEep:STEP.7 [MAXimumIMINimum]
Sets and queries the size of each frequency step. : STEP is governed by the equation
STEP = SPAN/POINTS.