Mtr Meas Menu

Function Group P O W E R

Menu Map 5

Description This softkey accesses the meter measure softkeys.

Meas C o r r A l l

Meas Corr current

Measures flatness correction values for all frequency points in the flatness correction array.

Measures a flatness correction value for the frequency point currently in the active line of the flatness correction array.

Meas Corr Undef Measures flatness correction values for all frequency points in the flatness correction array that have no correction values assigned.

The meter measure function uses an external HP 437B power meter to automatically measure and store power correction values for the frequency points requested.

Programming Codes

See Also

SCPI: NONE, see Fltness Menu

Analyzer: NONE

Flatness Menu

“Creating and Applying the User Flatness Array,” in Chapter 1.

M-28 Operating and Programming Reference

HP 8360

User’s

Handbook

Page 274
Image 274
HP 8360 manual Mtr Meas Menu, Function Group P O W E R Menu Map