ANALYZER STATUS REGISTER
STATUS BYTE (#l)
Bit # | 7 | 6 | 5 | 4 | 3 | 2 | 1 | 0 |
Decimal | 128 | 64 | 32 | 16 | 8 | 4 | 2 | 1 |
Value |
|
|
|
|
|
|
|
|
Function | SRQ on new | REQUEST | SRQ on | SRQ on | SRQ on | SRQ on | SRQ on | SRQ on |
| frequencies | SERVICE | End of | RF Settled | Changed in | Numeric | Any Front | |
| or sweep | (RQS) | syntax error. | Sweep |
| Extended | Entry | Panel Key |
| time in |
|
|
|
| status | Completed | Pressed |
| effect. |
|
|
|
| Byte |
| |
|
|
|
|
|
|
| Front |
|
|
|
|
|
|
|
| Panel) |
|
|
| EXTENDED STATUS BYTE (#2) |
|
|
| |||
|
|
|
|
|
|
|
|
|
Bit # | 7 | 6 | 5 | 4 | 3 | 2 | 1 | 0 |
Decimal | 128 | 64 | 32 | 6 | a | 4 | 2 | 1 |
Value |
|
|
|
|
|
|
|
|
Function | Fault | RF Unleveled | Power | RF Unlocked | External | Oven | Over | Self Test |
| Indicator |
| Failure |
| Frequency | Cold | Modulation | Failed |
| On |
|
|
| Reference |
|
|
|
|
|
|
|
| Selected |
|
|
|
Status Byte 1
Bit 0: SRQ caused by a key closure on the front panel of the synthesizer (use the OM code to determine the front panel status).
Bit 1: SRQ caused by the completion of a numeric entry (use the OA code to determine the value of the numerical entry).
Bit 2: SRQ caused by a change in the extended status byte (status byte 2) affected by the
Bit 3: SRQ caused by the completion of phase locking and the settling of the RF source (use the OK code to determine the last lock frequency).
Bit 4: SRQ on
Bit 5: SRQ caused by
Bit 6: SERVICE REQUEST; by
instrument needs service from the controller when this bit is set true.
Bit 7: SRQ caused by a change in the coupled parameters (start frequency, center frequency, and sweep time). Use the OC code to determine the new values of the coupled parameters.
Status Byte 2 (Extended Status Byte)
Bit 0: Self test failed at power on or at Instrument Preset. This bit remains latched until this status byte has been read, or until cleared by the CS or CLEAR 719 commands.
Bit 1: Excessive amplitude modulation input.
| HP 8360 |
User’s | Handbook |