Center=Narker

Function Group MARKER

Menu Map 3

Description This softkey sets the center frequency of the sweep to the frequency of the most recently activated marker. Select any marker Ml . . .

M5, then select Center=Marker to change the center frequency of the sweep to that of the marker. The frequency span does not change unless the new sweep limits fall outside the frequency range of the synthesizer, in that case the synthesizer automatically scales the frequency span to be within the synthesizer’s operating frequency range.

Programming Codes SCPI:

MARKer[n][:FREQuency] ?

FREQuency:CENTer <freq from above> [freq suffix]

Analyzer: MC

See Also [w)

“Marker Operation” in Chapter 1.

Clear Fault

Function Group

Menu Map

Description

Programming Codes

SERVICE

6

This softkey clears all the latched fault status indicators.

SCPI: DIAGnostics:OUTPut:FAULts

The above command relays the fault information and clears all faults. Analyzer: NONE

See Also Fault Menu
C-2 Operating and Programming Reference

HP 8360

User’s

Handbook