Programming Codes SCPI:

CORRection:FLATness {<num>[freq suffix],<num>[DB]}2*801

The portion of the above command contained in { } can be entered from one to 801 times. This command creates the frequency-correction pair array similar to the front panel array. The correction entered is at the associated frequency and frequencies in between are determined by linear interpolation.

CORRection:FLATness?

This command queries the flatness array created with CORR:FLAT.

CORRection:ARRay[i] {<num>[DB]}1601*1601

The portion of the above command contained in {} must

be entered 1601 times. This array must contain 1601 evenly spaced correction values. This command creates the 1601-point correction set that has no equivalent front panel entry. If this command is used to enter flatness correction information the CORRection:SOURce command (described below) will be set to array. There is an array for the foreground state (i=O) and for the background state (i=l). If [‘I1 is not specified, the default is the foreground state (i=O).

CORRection:ARRay[i]?

This command queries the entire 1601-point correction set.

CORRection:SOURce[i] ARRayIFLATness

When the above command is set to flatness CORR:SOUR FLAT, the array chosen is the frequency-correction pair array. When the command is set to array CORR:SOUR ARR, the array chosen is the 1601 point correction set.

CORRection:SOURce[i]?

Queries the source of correction.

CORRection[:STATe] ON]OFF]l]O

Sets the switch on the user flatness correction feature. This is the same as pressing FLTNESS ON/OFF on the front panel.

CORRection:STATe?

Queries the condition of the internal switch.

CORRection:FLATness:POINts? [MAXimumJMINimum]

The above command returns information on how many frequency-correction pairs were entered using the CORR : FLAT command.