(SPAN)

Function Group

Menu Map

Description

FREQUENCY

2

This softkey lets you set a value for the frequency span in the center frequency/frequency span mode of swept frequency operation.

Press w), and use the entry area to enter the desired value. The synthesizer sweeps from the span below to above the center frequency.

Certain center frequency and frequency span combinations cause the synthesizer to limit the value entered. In general, any combination that causes the synthesizer to exceed its minimum or maximum specified frequency is limited.

Programming Codes SCPI:

FREQuency:SPAN <num>[freq suffix] or

MAXimum]MINimum]UP]DOWN

FREQuency:MODE SWEep

Analyzer: DF <num> [Hz]Kz]Mz]Gz]

See Also [CENTER),ETxq,(STOP)

“Center Frequency/Span Operation,” in Chapter 1.

Function Group FREQUENCY

Menu Map 2

Description This softkey activates swept frequency mode and makes the start frequency parameter the active function. Press CSTART) and use the entry area to enter the desired value. The start/stop frequency must be separated by at least 2 Hz in order to remain in the frequency sweep mode. If start=stop frequency then the zero span mode is entered.

HP 8360

Operating and Programming Reference S-61

User’s Handbook