Meter Adrs

Function Group SYSTEM

Menu Map 8

Description In cases where the synthesizer is capable of acting as a controller to an HP 437B power meter, this softkey enables you to set the programming address of the power meter. The address value can be set from 0 to 30, with the factory default address set at 13. The address value is stored in non-volatile memory.

Programming Codes SCPI: DIAGnostics:INSTrument:PMETer:ADDRess <num> Analyzer: NONE

See Also Adrss Menu

“Optimizing Synthesizer Performance,” in Chapter 1.

“INSTALLATION,” Chapter 3.

Meter On/Off AM

Function Group

Menu Map

Description

Programming Codes

See Also

INIOD)

4

This softkey (Option 002 only) lets you display the value of the depth of the externally-generated amplitude modulation.

SCPI: MEASure:AM?

Analyzer: NONE

ml, also see “AM” and “Modulation”.