•ABORt

Causes the sweep in progress to abort and reset. If INIT: CONT is ON it immediately restarts the sweep. The pending operation flag (driving *OPC, *WAI, and *OPC?) undergoes a transition once the sweep is reset.

•AM[:DEPTh] Qnun>[PCTlIMAXimumlMINimuml~num>DB

•AM[:DEPThl?[MAXimumlMINimum]

Sets and queries the percentage of AM when the SOURce is INTernal. If <num> is received with units of dB, the value is converted to percent by the equation:

[%] = 100 * (1 - lO**(-dB/20))

Valid ranges of dB are 0 to 40 dB. After “RST, the value is 50%.

•AM:INTernal:FREQuency Qnun>[Cfreqsuffix>]lMAXimumlMINimum

•AM:INTernal:FREQuency?[MAXimumIMINimum]

Sets and queries the frequency (in Hz) of the internal AM source.

After *RST, the value is 1 kHz.

•AM:INTernal:FUNCtion SINusoidlSqUarelTRIanglelRAMPlNOISe

•AM:INTernal:FUNCtion?

Sets and queries the waveform of the internal AM source. After “RST, the value is SINusoid.

•AM:SOURce INTernallEXTernal

•AM:SOURce?

Sets and queries the source of the AM modulating signal. After *RST, the value is EXTernal.

•AM:MODE DEEPlNORMal a AM:MODE?

Controls the AM depth limits of the synthesizer. The NORMal position is selected at *RST.

•AM:STATe ONlOFFlllO

•AM:STATe?

Sets and queries the status of the AM modulation. After *RST, the

setting is OFF.

•AM:TYPE LINearlEXPonential a AM:TYPE?

Sets and queries the type of AM modulation. After *RST, the setting is LINear.

•CALibration:AM:AUTO ONlOFFlllO

a CALibration:AM:AUTO?

Sets and queries the automatic modulator calibration switch.

S-28 Operating and Programming Reference

HP 8360

User’s

Handbook

Page 326
Image 326
HP 8360 manual Sets and queries the automatic modulator calibration switch