•DIAGnostics:TEST:LOG[:STATe]?

•DIAGnostics:TEST:LOG[:STATe] ONOFFlO

Selects and queries the raw data logging ON/OFF switch. Both

commands are executable in selftest mode.

After *RST, the setting is 0.

•DIAGnostics:TEST:LOOP ONOFFlO

•DIAGnostics:TEST:LOOP?

Selects and queries the test looping ON/OFF switch. Both commands are executable in selftest mode.

After “RST, the setting is 0.

•DIAGnostics:TEST:NAME? [<num>]

Queries the name of a selftest by number. If the number is not specified then an array of all the selftest names is returned.

•DIAGnostics:TEST:POINts?

Returns the number of points of selftest that is output using

DIAGnostics:TEST:NAME? or DIAGnostics:TEST:RESult?.

•DIAGnostics:TEST:RESult? [<num>]

Queries the result of a selftest, by number. The response is a string containing either Passed, Failed, or NotRun. If <num> is missing, an array of selftest results are returned.

•DIAGnostics:TINT? <num>

A test feature that returns the value passed to it. This is used to test the HP-IB interface.

•DISPlay[:STATel ONOFFlO

•DISPlay[:STATe]?

Sets and queries the display ON/OFF switch.

After *RST, the value is 1.

•FM:COUPling ACiDC

•FM:COUPling?

Sets and queries the FM input coupling mode.

The *RST value is AC.

•FM[:DEViation] <num>[freq suffix>llMAXimumlMINimum

•FM[:DEViation]? [MAXimumlMINimum]

Sets and queries the peak FM deviation (in Hz) when the internal

FM generator is used. After 'RST, the value is 1 MHz.

•FM:FILTer:HPASs <num>[freq suffix'lMAXimumlMINimum

•FM:FILTer:HPASs? [MAXimumlMINimuml

Sets and queries the FM AC bandwidth. There are only two

S-34 Operating and Programming Reference

HP 8360

User’s

Handbook

Page 332
Image 332
HP 8360 manual Selects and queries the raw data logging ON/OFF switch. Both