Power Sweep

in dB/Hz, and t is either “DB” or the ASCII LF terminator. For example, for a slope of 1.5 dB/GHz use this procedure:

1.1.5 dB/GHz = 1.5 dB/1,000,000,000 Hz

2.1.5 dB/lES Hz = 1.5E-9 dB/Hz

3.The programming code is “SL11.5E-9 DB”

See Also CPOWER LEVEL), Power Sweep

“Power Sweep and Power Slope Operation” in Chapter 1.

Power Sweep

Function Group

Menu Map

Description

POWER

This softkey enables the power sweep function. RF output power can be swept both positively and negatively over a selected range. The level of the power sweep starting point is the power level programmed. Power sweep widths can be 45 dB wide in either direction. However, the settable power sweep range is dependent on the ALC level set. An asterisk next to the key label indicates that this feature is active.

Programming Codes

See Also

SCPI:

POWer:MODE SWEeplFIXed

POWer:STARt <num>[level suffix]]MAXimum]MINimum POWer:SPAN <num>[level suffix]]MAXimum]MINimum

Analyzer: PSl function on, PSO function off.

POWER LEVEL , Power Slope

“Power Sweep and Power Slope Operation” in Chapter 1.

HP 8360

Operating and Programming Reference P-7