Note | U, V, and |
|
| ||
| Power Meter with the | |
| (50 to 75 GHz), or the MP81B (75 to 110 GHz) power sensors. Since | |
| the Anritsu model ML83A Power Meter is not capable of internally | |
| storing power sensor cal factors, you must manually correct the | |
| data entry. Refer to example 2 for information on manual entry of | |
| correction | data. |
|
|
|
Setup Synthesizer Parameters
5.Turn on the synthesizer and press [PRESET). The following occurs:
n The source module’s frequency span is displayed on the synthesizer.
n The synthesizer’s leveling mode is automatically changed from internal to “module leveling”.
n The source module’s maximum specified power is set and displayed.
6.Press FREQUENCY (START) @ @ 0 @ (GHz, m@ @ IGHz). The frequency sweep is set from 26.5 to 40 GHz.
7.Press (POWER LEVEL) 0 @. The source module power is set to +7 dBm for maximum power to the device under test.
Access User Flatness Correction Menu
8.Press POWER CE). Select Fltness Menu.
9.Select Delete Menu Delete All. This step insures that the flatness array is empty.
10.Press (PRIOR). Leave the delete menu and return to the previous softkey menu.
11.Select Enter Freq (!?J @ 0 @ (GHz), to enter 26.5 GHz as the first correction frequency. Enter 31, 32.5, and 40 GHz to complete the list. Notice that the frequencies are arbitrarily spaced.
Enter Correction Data into Array
12.Select Mtr Meas Menu Measure Corr All. The power meter is now under synthesizer control and is performing the sequence of steps necessary to generate the correction information at each frequency point.
If an