SCPI COMMAND SUMMARY
Sets and queries the save/recall register number to alternate the foreground state of the instrument. The *RST value is 1.
•SYSTem:ALTernate:STATe ONlOFFlllO
•SYSTem:ALTernate:STATe?
Sets and queries the state of the Alternate State function. *RST setting is OFF.
•SYSTem:COMMunicate:GPIB:ADDRess <n>
Changes the GPIB’s (General Purpose Interface Bus) address.
The *RST value is 19.
•SYSTem:DUMP:PRINter?
Causes a dump of the display contents to be made to the
•SYSTem : ERRor?
Returns the next message in the error queue. The format of the response is : <error number>,<error string>
where the error number is as shown in the “Error Messages” section and error string is :
“<Generic
An example response to SYST:ERR? is
-23, “NUMERIC OVERFLOW ;YOU PUT IN A NUMBER TOO BIG”
•SYSTem:KEY:ASSign <num>,<num>
This assigns the first numeric value (key code) to the second numeric value (user menu key number). Every softkey menu item is given a unique key code (as if it were a dedicated front panel key). Refer to the entry SCPI KEY NUMBERS for a listing of key codes.
•SYSTem:KEY:CLEar <num>lALL
Clears the user menu. If
•SYSTem:KEY :DISable SAVE
The save key grouping is disabled. In our box this disables the save state feature. (Save Lock)
a SYSTem:KEY:ENABle SAVE
Unlocks the save registers. The *RST value is for the save registers to be enabled.
•SYSTem:LANGuage SCPIlCIILlCOMPatible
Causes the synthesizer to perform a language switch to another language system.
HP 8360 | Operating and Programming Reference |