Sweep Mode Ramp

Programming Codes SCPI: NONE

Analyzer: NONE

See Also Softkeys listed above.

“Programming Typical Measurements,” in Chapter 1.

Sweep Mode List

Function Group

Menu Map

Description

SWEEP

7

This softkey activates the step frequency list mode. To use this type of sweep, a frequency list must have been entered, otherwise an error message appears. In this mode, the synthesizer steps only those frequencies defined by the frequency list. An asterisk next to the key label indicates that this feature is active.

Programming Codes SCPI: FREQuency:MODE LIST

Analyzer: SN

See Also CONNECTORS, List Menu

“Creating and Using a Frequency List,” in Chapter 1.

Sweep Mode Ramp

Function Group SWEEP

Menu Map 7

Description This softkey activates the analog frequency sweep mode. Ramp sweep mode is the factory preset state. An asterisk next to the key label indicates that this feature is active.

HP 8360

Operating and Programming Reference S-73