Pulse Rise Time SlQW

Function Group

Menu Map

Description

Programming Codes

See Also

MODULATION

4

This softkey lets you set the synthesizer to apply a slow rise pulse filter to both internal and external pulse waveforms. This results in pulses having approximately 2 ps rise/fall times. An asterisk next to the key label indicates that this function is active.

SCPI: PULM:SLEW <num>[time suffix]MAXimum

Analyzer: NONE

Pulse Menu

Pulse Width

Function Group

Menu Map

Description

Programming Codes

See Also

(MOD)

4

This softkey lets you set a value for the internal pulse generator’s pulse width. The range of acceptable values is from 1 /.JS to 65.5 ms. The factory preset value is 1 ms. When this feature is active, its current value is displayed in the active entry area.

SCPI: PULSe:INTernal:WIDTh <num>[time suffix] or MAXimumIMINimum

Analyzer: NONE

m), m, Pulse Menu

P-22 Operating and Programming Reference

HP 8360

 

User’s Handbook

Page 296
Image 296
HP 8360 manual Pulse Rise Time SlQW, Pulse Width