[POWER LEVEL)
MTR: x.xx .
In Normal, Module, the (POWER LEVEL) key controls the output power of the synthesizer as compared to the feedback voltage from
a
Figure A-l) within the ALC level range (+25 to -20 dBm). This mode of operation requires a feedback connection from the module to the synthesizer through the SOURCE MODULE INTERFACE.
When you press [POWER LEVEL), the active entry area displays:
where X represents a numeric value. The data display area indicates:
Power(dBm)
MDL: x.xx .
In ALCoff, there is no feedback voltage to level the power, so power level is uncalibrated. A leveling point is not specified in this mode. The (POWER LEVEL) key controls the linear modulator directly, from 0 to approximately
When you press [POWER LEVEL), the active entry area displays:
where X represents a numeric value. The data display area indicates:
Ref(dB)
OFF: x.xx
and the message line indicates: UNLVLED.
In Search, any of the leveling points can be specified and used as the comparison feedback voltage. Basically, this mode operates the same as ALCoff after the
Programming Codes SCPI: POWer[:LEVEL] <num>[lvl suffix] or
MAXimum]MINimum]UP]DOWN
Analyzer: PL
See Also (ALC), CONNECTORS, Det Gal Menu, (FLTNESS ON/OFF],
Set Atten, Tracking Menu, Uncoupl Atten
“Programming Typical Measurements,” in Chapter 1.
| HP 8360 |
| User’s Handbook |