Table 1-4. Sample Synthesizer Commands (continued)

Command

Parameters

Parameter Type

Allowed Values

 

 

 

 

 

POWer

 

 

 

 

:ATTenuation

atten setting

extended numeric

0 to 90 [DB] or

 

 

 

 

MAXimum]MINimum/UPlDOWN

:AUTO

coupled atten

Boolean

 

ONjOFF]l/O

[:LEVel]

output level

extended

numeric

specified power range

 

 

 

 

Or

 

 

 

 

MAXimum]MINimum]UP]DOWN

:STATe

RF on/off

Boolean

 

ON]OFF]l]O

 

 

 

 

 

SWEep

 

 

 

 

:GENeration

type of sweep

discrete

 

STEPpedlANALog

:TIME

sweep time

extended

numeric

200s to 133 ms or

 

 

 

 

MAXimumlMINimum

:AUTO

auto sweep

Boolean

 

ON]OFF]l]O

 

time switch

 

 

 

:LLIMit

fastest sweep

extended numeric

<num>[time suffix] or

 

time

 

 

MAXimumlMINimum

 

 

 

 

 

HP-IB Chec Example lrogram 1

This first program is to verify that the HP-IB connections and interface are functional. Connect a controller to the synthesizer via an HP-IB cable. Clear and reset the controller and type in the following program:

10Source=719

20ABORT 7

30LOCAL Source

40CLEAR Source

50REMOTE Source

60CLS

70PRINT "The source should now be in REMOTE."

80PRINT "Verify that the 'REMOTE' LED is on."

90END

Run the program and verify that the REMOTE LED is lit on the synthesizer. If it is not, verify that the synthesizer address is set to 19 and that the interface cable is properly connected.

If the controller display indicates an error message, it is possible that the program was entered in incorrectly. If the controller accepts the REMOTE statement but the synthesizer REMOTE LED does not turn on, perform the operational checks as outlined in the respective Operating and Service Manuals to find the defective device.

1-88 Getting Started Programming

Page 118
Image 118
HP 8360 manual HP-IB Chec Example lrogram, Command Parameters Parameter Type Allowed Values POWer