Ml--M2 Sweep

Function Group

Menu Map

Description

Programming Codes

See Also

MARKER

3

This softkey lets you set the synthesizer to start sweeping at the frequency of marker 1 (Ml), and stop sweeping at the frequency of marker 2 (M2). M2 must have a higher frequency value than Ml. If Ml--M2 Sweep is activated when M2 is at a lower frequency than Ml, the values of Ml and M2 are permanently interchanged. While this function is active, the start/stop frequencies of the synthesizer are changed to the values of Ml and M2. An asterisk next to the key label indicates this feature is active.

SCPI: SWEep:MARKer:STATe ONIOFFJ110

Analyzer: MPl function on, MPO function off.

Marker Ml, Start=Ml Stop=MZ

“Marker Operation,” in Chapter 1.

“Programming Typical Measurements,” in Chapter 1.

Manual Sweep

Function Group SWEEP

Menu Map 7

Description This softkey lets you set the synthesizer to the manual sweep mode of operation. Depending on what parameter is sweeping, you can use either the rotary knob or the ARROW keys to manually sweep between the start/stop limits. In manual sweep mode, the synthesizer does not automatically retrace at the sweep end point (the user must turn the rotary knob to retrace), and the green SWEEP LED does not light. The resolution of the rotary knob is 0.1% of the sweep span in either start/stop or CF/AF mode. The resolution of the

a and a arrow keys are dependent on the resolution defined by

HP 8360

Operating and Programming Reference M-l

User’s Handbook

 