HP 8360
User’s Handbook
SWEep : STEP value is
SWEep:TIMEMINimum
SWEep:TIME:AUTO ON
SWEep:TIME:LLIMit 10ms
SWEep : GENerat ion ANALog
SWEep : MODE AUTO
SWEep : MANual : POINt 1
SWEep:MANual[:RELative] 0.50
SWEep:MARKer:STATe OFF
SYSTem: ALTernate 1
SYSTem:ALTernate:STATe OFF
SYSTem:COMMunicate:GPIB:ADDRess 19
SYSTem:KEY:ENABle SAVE
SYSTem:MMHead:SELect:AUTO ON
SYSTem:SECurity:COUNt 1
UNIT:AMPCT *UNIT:POWerDBM
•*SAV <num>
The present instrument state is stored in the specified memory register. The acceptable numeric range is from 1 to 8. An execution error occurs if you try to save state 0.
•*SRE <num> • *SRE?
Sets and queries the value of the Service Request Enable Register.
•*STB?
Queries the Status Byte. This is a
•*TRG
This command performs the same function as the Group Execute Trigger command defined by IEEE 488.1.
•*TST?
A full selftest is performed, without data logging or looping, and returns one of the following error codes:
Error Code Definition
0Test passed.
1Test failed.
2Test not run yet. (This is an unlikely event.)
3Test aborted.
4Can not execute the test.
5Can not execute the test, test skipped.
a *WA1
This causes the synthesizer to wait for the pending commands to be completed before executing any other commands. For example,