Appendix B: Performance Verification (AWG710B)
B-22 AWG710&AWG710B Arbitrary Waveform Generator User Manual
d. Push 0, ., 0, 2 and ENTER keys in this order or turn the general purpose
knob to set the amplitude to 0.020 V.
3. Push the RUN and CH1 output buttons.
The LEDs above the RUN button and CH1 output connector are on.
4. Do the following substeps to set the AWG710B Arbitrary Waveform
Generator offset:
a. Push the Offset side button.
b. Push 0 and ENTER keys in this order.
c. Verify that the reading on the DMM falls within 0 V ± 10 mV.
d. Push 0, ., 5 and ENTER keys in this order to change the AWG710B
Arbitrary Waveform Generator offset to 0.5 V.
e. Verify that the reading on the DMM falls within 0.5 V ± 0.0175 V.
f. Push , 0, ., 5 and ENTER keys in this order to change the AWG710B
Arbitrary Waveform Generator offset to –0.5 V.
g. Check that the reading on the DMM falls within –0.5 V ± 0.0175 V.
5. Do the following substeps to change the connection to check the CH1:
a. Push the CH1 output button.
b. Disconnect the adapter, terminator and cable from the CH1 output
connector.
c. Connect the adapter, terminator, and cable (removed in st ep b) to the CH1
output connector.
d. Push the CH1 output button to turn on the CH1 LED.
6. Repeat the Check Offset Accuracy procedure for the AWG710B Arbitrary
Waveform Generator CH1.
7. Push the CH1 output button to turn off the CH1 LED.
8. Disconnect the test hookup from the CH1 output connector.