Appendix B: Performance Verification (AWG710)
AWG710&AWG710B Arbitrary Waveform Generator User Manual B-89
Figure B-50: Trigger signal and waveform output (+5 V check 1)
e. Change the pulse waveform high level to 5.35 V by using cursor keys.
f. Check that sine waves are displayed on the oscilloscope as sho wn in Figure
B-41.
Figure B-51: Trigger signal and waveform output (+5 V check 2)
7. Follow the substeps below to check trigger level accuracy at –5 V:
a. Push –, 5 and ENTER keys in this order.
b. Set the DC waveform low level of function generat or to –5.35 V by using
cursor keys.
Trigger
signal
Waveform
output CH1
CH2
Tri g g e r
signal
Waveform
output CH1
CH2