Appendix B: Performance Verification (AWG710)
B-100 AWG710&AWG710B Arbitrary Waveform Generator User Manual
Marker Output Tests
These procedures check the accuracy of the AWG710 Arbitrary Waveform
Generator marker output level.
Do the following steps to install the test ho okup and set the t est equipment controls:
1. Use a 50 BNC coaxial cable, a SMA(Ma)–BNC(Fe) adapter, a 50
precision terminator, and a BNC–to–dual banana adapter to connect the
AWG710 Arbitrary Waveform Generator MARKER 1 OUT to the DMM
input connector (see Figure B-63).
Figure B-63: Marker output initial test hookup
2. Set the DMM controls as follows:
NOTE. Connect a 50

SMA terminator to the inverted marker output connector
during the marker output tests.
Equipment
required
A 50 BNC coaxial cable, a 50 precision terminator, a
SMA(Ma)–BNC(Fe) adapters, BNC (female)–to–dual banana adapter,
and a digital multimeter (DMM).
Prerequisites The AWG710 Arbitrary Waveform Generator must meet the prerequisites
listed on page B-61.
AWG710 Arbitrary Waveform Generator
DMM
50 BNC coaxial cable
MARKER 1 out connector
+ SMA(Male)-BNC(Female) adapter
+ BNC 50 precision terminator
+ 50 BNC coaxial cable
Input connector
+ BNC-to-dual banana adapter
+ 50 BNC coaxial cable