Appendix B: Performance Verification (AWG710)
AWG710&AWG710B Arbitrary Waveform Generator User Manual B-99
b. Set the function generator (AFG310) controls:
Check 1/4 Clock
frequency 1. Follow the substeps below to set t he AWG710 Arbitrary Waveform Generator
controls and select the waveform file:
a. Push UTILITY (front–panel)!System (bottom)!Factory Reset
(side)!OK (side).
b. Load the MODE.WFM file.
Refer to Loading Files on page B-62 for file loading procedures.
c. Push HORIZONTAL MENU (front–panel)!Clock (side).
d. Push 2, 0, 0 and M (SHIFT+7) keys in this order or turn the genera l
purpose knob to set the internal clock frequency to 200 MHz.
2. Push the RUN button.
The LED above the RUN button is on.
3. Verify that the frequency counter reading is 50 MHz ± 50 Hz (using internal
reference clock).
4. Retain the test hookup.
Check 10MHz Reference
Input 5. Push HORIZONTAL MENU (front–panel)!Clock Ref (side) so that the
AWG710 Arbitrary Waveform Generator reference clock is set to External.
6. Verify that the fr equency coun ter reading is 50.0 MHz ±10 kHz (usi ng external
reference clock).
7. Change the function generator controls as follows:
8. Check that the frequency counter reading is 50.5 MHz ±10 kHz.
9. Turn the function generator output off and disconnect the function generator
and frequency counter.
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Square
Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous
Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 MHz
Amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.0 V into 50 (4.0 V into 1 M)
Offset. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 V
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . On
Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.1 MHz