![](/images/new-backgrounds/1303623/303623205x1.webp)
MODEL 3081 pH/ORP | SECTION 12.0 |
| TROUBLESHOOTING |
SECTION 12.0
TROUBLESHOOTING
12.1WARNING AND FAULT MESSAGES
12.2CALIBRATION ERRORS
12.3TROUBLESHOOTING - GENERAL
12.4TROUBLESHOOTING WHEN A DIAGNOSTIC MESSAGE IS SHOWING
12.5TROUBLESHOOTING WHEN NO DIAGNOSTIC MESSAGE IS SHOWING
12.6SYSTEMATIC TROUBLESHOOTING
12.7DISPLAYING DIAGNOSTIC VARIABLES
12.8TESTING THE TRANSMITTER BY SIMULATING pH
12.9FACTORY ASSISTANCE AND REPAIRS
12.1 WARNING AND FAULT MESSAGES
The Model 3081 pH/ORP transmitter continuously monitors the measurement loop (sensor and transmitter) for conditions that cause erroneous measurements. When a problem occurs, the transmitter displays either a warning or fault mes- sage. A warning alerts the user that a potentially system disabling condition exists. If the condition causing the problem is not corrected, there is a high probability that the system will soon fail. A fault alerts the user that a system disabling con- dition exists. If a fault message is showing, all measurements should be regard- ed as erroneous.
When a WARNING condition exists:
1.The main display remains stable; it does not flash.
2.A warning message appears alternately with the temperature/output display. See Figure
When a FAULT exists:
1.The main display flashes.
2.The words FAULT and HOLD appear in the main display.
3.A fault message appears alternately with the temperature/output display. See Figure
4.The output current will remain at the present value or go to the programmed fault value. See Section 8.3 Output Ranging for pH Measurements, or Section 10.3 Output Ranging for ORP Measurements for details on how to program the current generated during a fault condition.
5.If the transmitter is in HOLD when the fault occurs, the output remains at the programmed hold value. To alert the user that a fault exists, the word FAULT appears in the main display, and the display flashes. A fault or diagnostic message also appears.
6.If the transmitter is simulating an output current when the fault occurs, the transmitter continues to generate the simulated current. To alert the user that a fault exists, the word FAULT appears in the display, and the display flash- es.
FIGURE
Annunciation
When a
FIGURE 12-2. Fault Annunciation
When a disabling condition, a fault, occurs, the display appears as pic- tured above. To further alert the user that measurements are in error, the display flashes. Diagnostic mes- sages appear in the temper- ature/output area on the screen.
89