MODEL 3081 pH/ORP

SECTION 12.0

 

TROUBLESHOOTING

12.7 DISPLAYING DIAGNOSTIC VARIABLES

12.7.1 Purpose

This section describes how to display the diagnostic variables listed below:

DIAGNOSTIC MEASUREMENTS

1.Sensor voltage in mV (InPut)

2.Glass impedance in megohms (GIMP)

3.Reference impedance in kilohms* (rIMP)

4.Temperature in °C (tEMP)

*For high impedance reference electrodes, the reference impedance is in megohms.

DIAGNOSTIC MESSAGES

1.Software version (VEr)

2.Display last three fault messages (ShoW FLt)

For an explanation of the meaning of diagnostic messages, refer to Section 8.4. Displays are read only.

12.7.2 Procedure

1.Enter the Diagnostic menu by pressing DIAG on the IRC. Sensor voltage in mV (InPut) appears.

2.Press NEXT . The temperature corrected glass impedance in megohms (GIMP) appears.

3.Press NEXT . The reference impedance (rIMP) appears. For conventional low impedance silver/silver chloride reference electrodes, the reference impedance has units of kilohms. For the rare occasions when a high impedance reference is used, the units are megohms. See Sections 8.4 (for pH) and 10.4 (for ORP) for more information.

4.Press NEXT . The model number and software version (Ver) appears.

5.Press NEXT . The temperature (tEMP) measured by the sensor appears.

6.Press NEXT . The ShoW Flt sub-menu appears.

7.Press ENTER . The most recent fault message appears in the display. Press NEXT repeatedly to scroll through the stored messages. The transmitter only remembers the three most recent messages. nonE appears if there are no faults. Pressing EXIT clears all the stored messages and returns the transmitter to the ShoW Flt display. If the transmitter loses power, all stored warning and fault messages are lost.

8.Press EXIT to return to the process display.

12.8 TESTING THE TRANSMITTER BY SIMULATING THE pH.

12.8.1 General.

This section describes how to simulate a pH input into the 3081 pH/ORP transmitter. pH is directly proportional to voltage. To simulate the pH measurement, connect a standard millivolt source to the transmitter. If the transmitter is working prop- erly, it will accurately measure the input voltage and convert it to pH. Although the general procedure is the same, the wiring details depend on the location of the preamplifier. Consult the table to find the correct procedure.

Preamplifier located in

Section

 

Transmitter

12.8.2

 

Remote junction box

12.8.3

 

Sensor-mounted junction box

12.8.3

 

Sensor (Model 381+ only)

12.8.4

 

Sensor (all other models)

12.8.5

 

 

 

 

110

Page 124
Image 124
Emerson Process Management 3081 pH/ORP Testing the Transmitter by Simulating the pH, Diagnostic Measurements