MODEL 3081 pH/ORP

 

 

 

SECTION 3.0

 

 

 

 

 

WIRING

Table 3-7 Wiring Diagrams for Model 385+ Sensors

 

 

 

 

 

 

 

 

 

Sensor

Junction Box

Preamplifier

RTD

Wire Functions

Wiring Diagram

385+ -02

sensor-mounted

in sensor-mounted junction box

Pt 100

Figure 3-15

Figure 3-16

 

 

 

 

 

 

385+ -03

none

in sensor

Pt 100

Figure 3-13

Figure 3-14

385+ -03

remote

in sensor

Pt 100

Figure 3-13

Figure 3-14

385+ -04

none

in transmitter

Pt 100

Figure 3-10

Figure 3-11

385+ -04

remote

in remote junction box

Pt 100

Figure 3-10

Figure 3-12

Table 3-8 Wiring Diagrams for Model 381+ Sensors

Sensor

Junction Box

Preamplifier

RTD

Wire Functions

Wiring Diagram

381+ -40-55

none

in sensor

Pt 100

Figure 3-13

Figure 3-14

 

 

 

 

 

 

381+ -43-55

none

in sensor

Pt 100

Figure 3-13

Figure 3-14

 

 

 

 

 

 

381+ -40-55

remote

in sensor

Pt 100

Figure 3-13

Figure 3-14

 

 

 

 

 

 

381+ -43-55

remote

in sensor

Pt 100

Figure 3-13

Figure 3-14

 

 

 

 

 

 

381+ -41-52

none

in transmitter

Pt 100

Figure 3-10

Figure 3-11

381+ -41-52

remote

in remote junction box

Pt 100

Figure 3-10

Figure 3-12

Table 3-9 Wiring Diagrams for Model 381pHE and 381pH Sensors

Sensor

Junction Box

Preamplifier

RTD

Wire Functions

Wiring Diagram

 

 

 

 

 

 

381pH-30-41-52*

none

in transmitter

3K Balco**

Figure 3-2

Figure 3-4

 

 

 

 

 

 

381pH-30-41-52*

remote

in remote junction box

3K Balco**

Figure 3-2

Figure 3-5

 

 

 

 

 

 

381pH-30-42-52

none

in transmitter

3K Balco**

Figure 3-3

Figure 3-4

 

 

 

 

 

 

381pH-30-42-52

remote

in remote junction box

3K Balco**

Figure 3-3

Figure 3-5

 

 

 

 

 

 

381pHE-31-41-52*

none

in transmitter

Pt 100

Figure 3-2

Figure 3-4

 

 

 

 

 

 

381pHE-31-41-52*

remote

in remote junction box

Pt 100

Figure 3-2

Figure 3-5

 

 

 

 

 

 

381pHE-31-42-52

none

in transmitter

Pt 100

Figure 3-3

Figure 3-4

 

 

 

 

 

 

381pHE-31-42-52

remote

in remote junction box

Pt 100

Figure 3-3

Figure 3-5

Table 3-10 Wiring Diagrams for Model 328A Sensor

Sensor

Junction Box

Preamplifier

RTD

Wire Functions

Wiring Diagram

328A

none

in transmitter

none

Figure 3-17

Figure 3-18

Table 3-11 Wiring Diagrams for Model 320HP Sensor

 

 

 

 

 

 

 

 

 

 

 

Sensor

 

Junction Box

Preamplifier

 

RTD

Wiring Diagram

320HP-10-55

 

on mounting plate

in transmitter

 

Pt 100

Figure 3-19

 

 

 

 

 

 

320HP-10-58

 

on mounting plate

in junction box attached to mounting plate

Pt 100

Figure 3-20

*Sensors have a BNC connector that the Model 3081 pH/ORP transmitter does not accept. Cut off the BNC and terminate the coaxi- al cable as shown in Figure 3-23. Alternatively, use a BNC adapter (PN 9120531).

**Set the RTD jumper to the 3K position (see Section 2.2). Also, program the transmitter to recognize the 3K RTD (see Section 8.5 for pH or 10.5 for ORP).

17

Page 31
Image 31
Emerson Process Management 3081 pH/ORP Wiring Diagrams for Model 385+ Sensors Junction Box, Wire Functions