MODEL 3081 pH/ORP | SECTION 3.0 |
| WIRING |
REMOVE BNC AND TERMINATE COAXIAL CABLE BEFORE WIRING SENSOR TO TRANSMITTER. SEE FIGURE
FIGURE 3-6. Wire functions for Models 397-50, 397-54, 396-50,
396-54, 396R-50-60, 396R-54-60, 389-02-50, and 389-02-54
before removing BNC and terminating cable.
IF USING A BNC ADAPTER, THE RED WIRE IS MV OR pH IN AND THE BLACK WIRE IS REFERENCE IN. TO PREVENT SHORT CIRCUITS TO THE TRANSMITTER HOUSING, INSULATE THE BNC WITH BY WRAPPING IT WITH ELECTRICAL TAPE.
FIGURE 3-7. Wire functions for Models 397-50, 397-54, 396-50, 396-54, 396R-50-60, 396R-54-60, 389-02-50, and 389-02-54 after removing BNC and terminating cable. Wire functions for Models 397-54-62, 396-54-62, and 389-02-54-62 as received.
3
NOTES:
1.PLACE PREAMPLIFIER SELECTION SWITCH S1 IN "TRANSMITTER" POSITION (SEE SECTION 2.2).
2.IF SENSOR HAS 3K BALCO RTD, SET JUMPER (SECTION 2.2) AND PROGRAM TRANSMITTER TO RECOGNIZE RTD (SECTION
3.JUMPERS SUPPLIED BY CUSTOMER.
FIGURE 3-8. Wiring diagram for Models 397-50, 397-54, 396-50, 396-54, 389-02-50, and 389-02-54 after removing BNC and termi- nating cable. Wiring diagram for Models 397-54-62, 396-54-62, and 389-02-54-62 as received. Wiring directly to the transmitter.
4
NOTES:
1.PLACE PREAMPLIFIER SELECTION SWITCH S1 IN "SEN- SOR/JUNCTION BOX" POSITION (SEE SECTION 2.2).
2.IF SENSOR HAS 3K BALCO RTD, SET JUMPER (SECTION 2.2) AND PROGRAM TRANSMITTER TO RECOGNIZE RTD (SECTION
3.SEE FIGURE
4.JUMPERS SUPPLIED BY CUSTOMER.
FIGURE 3-9. Wiring diagram for Models 397-50, 397-54, 396-50, 396R-50-60, 396R-54-60, 396-54, 389-02-50, and 389-02-54 after removing BNC and terminating cable. Wiring diagram for Models 397-54-62, 396-54-62, and 389-02-54-62 as received. Wiring through a remote junction box to the transmitter.
19