MODEL 3081 pH/ORP | SECTION 14.0 |
| ORP MEASUREMENTS |
The ORP of the iron (II) - iron (III) standard when meas- ured with a platinum electrode against a saturated sil-
Quinhydrone - hydroquinone ORP standards are also used. They are prepared by dissolving excess quinhy- drone in either pH 4.00 or pH 6.86 buffer. The ORP of the standards at a platinum electrode against a silver - silver chloride reference has been measured at 20°C, 25°C, and 30°C.
Temperature | ORP in | ORP in |
| pH 4.00 buffer | pH 6.86 buffer |
|
|
|
20°C | 268 mV | 92 mV |
|
|
|
25°C | 263 mV | 86 mV |
|
|
|
30°C | 258 mV | 79 mV |
There are two disadvantages to using quinhydrone standards. First, the shelf life is only about eight hours, so fresh standard must be prepared daily. Second, hydroquinone is highly toxic, so preparing, handling, and disposing of the standards requires care.
Unlike pH calibrations, which are generally done using two calibration buffers, ORP calibrations are almost always single point calibrations.