MODEL 3081 pH/ORP

SECTION 8.0

 

PROGRAMMING FOR pH MEASUREMENTS

PROGRAM

OutPut

EXIT NEXT ENTER

8.3.3 Procedure

1. Enter the Program menu by pressing PROG on the IRC. The OutPut sub-menu appears.

PROGRAM

4MA

 

0

0.00

EXIT

 

 

 

 

 

ENTER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PROGRAM

 

 

 

 

 

 

20MA

1

 

4.00

 

EXIT

 

 

 

 

 

ENTER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PROGRAM

 

 

 

 

 

 

HoLd

 

 

 

 

1 . 0 0

2

EXIT

 

 

 

 

 

ENTER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PROGRAM

 

 

 

 

 

 

FAULt

 

 

 

 

2.00

2

EXIT

 

 

 

 

 

ENTER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PROGRAM

 

 

 

 

 

 

dPn

 

00

0

EXIT

 

 

 

 

 

ENTER

 

 

 

 

 

 

 

2. Press ENTER . The screen displays the 4 MA prompt. Use the editing keys to change the displayed number to the desired pH. The allowed range is 0.00 to 14.00. Press ENTER to save.

3. The screen displays the 20 MA prompt. Use the editing keys to change the displayed number to the desired pH. The allowed range is 0.00 to 14.00. Press ENTER to save.

4. The screen displays the HoLd prompt. Use the editing keys to change the display to the output desired when the transmitter is in hold. The range is 3.80 to 22.00 mA. Entering 00.00 causes the transmitter to hold the output at the value it was when placed in hold. The hold setting overrides the fault setting. Press ENTER to save.

5.The screen displays the FAULt prompt. Use the editing keys to change the display to the output desired when the transmitter detects a fault. The range is 3.80 to 22.00 mA. Entering 00.00 causes the transmitter to hold the output at the value it was when the fault occurred. Press ENTER to save.

6. The screen displays the dPn prompt. Use the editing keys to change the display to the

desired output dampening. The range is 0 to 255. Press ENTER to save.

7. Press EXIT twice to return to the process display.

59

Page 73
Image 73