5.2 INFRARED REMOTE CONTROLLER (IRC) - KEY FUNCTIONS
The infrared remote controller is used to calibrate and program the transmitter and to read diagnostic messages. See Figure
RESET - Press to end the current oper- ation and return to the process display. Changes will NOT be saved. RESET does not return the transmitter to factory default settings.
Editing Keys - Use the editing keys to change the value of a flashing display. The left and right arrow keys move the cursor one digit at a time across a num- ber. The up and down arrow keys increase or decrease the value of the selected digit. The up and down arrow keys also scroll the display through the items in a list.
CAL - Press to access the calibrate menu.*
PROG - Press to access the program menu.*
DIAG - Press to view diagnostic mes- sages.*
*Pressing CAL, PROG, or DIAG causes the program screen to appear with the selected menu (CALIBRATE, PROGRAM, OR DIAGNOSE) showing. See Figure
HOLD - Press to access the prompt that
turns on or off the Hold function.
ENTER - Press to advance from a sub- menu to the first prompt under the sub- menu. Pressing ENTER also stores the selected item or value in memory and advances to the next prompt.
NEXT - Press to advance to the next
EXIT - Press to end the current opera- tion. The transmitter returns to the first prompt in the present
FIGURE 5-3. Infrared Remote Controller.
Hold the IRC within 6 feet of the transmitter, and not more than 15 degrees from the center of the display window.