MODEL 3081 pH/ORP

SECTION 12.0

 

TROUBLESHOOTING

12.4.4 rEFWArn

rEF WArn is an electrode fault message. It means the reference electrode impedance exceeds the programmed Reference Warning High (RWH) limit. Ideally, when the measurement system exceeds the warning limits, the user will have adequate time to diagnose and correct problems before a failure occurs. A high reference impedance implies that the liquid junction is plugged or the reference electrolyte is depleted. The message also appears if an inappropriate limit has been entered into the transmitter.

If the measurement system was previously commissioned and operating correctly, rEF WArn likely means a real problem exists. However, if the system is being started up or if the advanced diagnostic feature is being used for the first time, rEF WArn could be caused by a miswired sensor or by programmed limits that are not correct for the sensor.

NOTE

rEF WArn is a sensor diagnostic message. Sensor diagnostic messages are optional. They can be turned off. To disable sensor diagnostic mes- sages, refer to Section 8.4.3.

Troubleshooting Flowchart - rEF WArn

Troubleshooting rEF WArn problems is exactly the same as troubleshooting rEF FAIL problems. Refer to Section 12.4.3.

12.4.5 CALIbrAtE

CALIbrAtE is a diagnostic intended for future use. If the CALIbrAtE message is showing go to Section 8.4 and disable CALIbrAte.

12.4.6 tEMP HI and tEMP LO

tEMP HI and tEMP LO mean the transmitter has detected a problem with the temperature measuring circuit. The problem may lie in the sensor, the cable, or the transmitter. The determination of temperature is an integral part of the pH meas- urement. Therefore, failure of the temperature measuring circuit is a system disabling condition. However, in an emer- gency, automatic temperature compensation can be disabled and the transmitter placed in manual temperature compensation. Refer to Section 8.5. For manual temperature compensation, choose a temperature equal to the average temperature of the process. The resulting pH reading will be in error. The more variable the temperature and the further the pH from 7, the greater the error.

Troubleshooting Flowchart- tEMP HI and tEMP LO

A.Check wiring, jumper settings, and software settings.

1.Check the wiring between the sensor and the transmitter. Refer to the appropriate wiring diagram in Section 3.0. Pay particular attention to TB-3 (RTD RTN), TB-4 (RTD SN), and TB-5 (RTD RTN). (NOTE: TB-3 means terminal 3 on the terminal board.)

2.Be sure the position of the RTD jumper on the analog board matches the type of RTD in the sensor. See Section 2.2.

3.Be sure the software settings in Section 8.5 match the type of RTD in the sensor.

If the diagnostic message disappears, the sensor is in good condition.

If the message persists, go to step B.

95

Page 109
Image 109
Emerson Process Management 3081 pH/ORP REFWArn, Troubleshooting Flowchart rEF WArn, CALIbrAtE, TEMP HI and tEMP LO