MODEL 3081 pH/ORP

SECTION 8.0

 

PROGRAMMING FOR pH MEASUREMENTS

TABLE 8-1. pH Settings List

ITEM

MNEMONIC

DISPLAY LIMITS

FACTORY SETTINGS

USER SETTINGS

PROGRAM LEVEL (Sections 8.0 - 8.9).

 

 

 

 

 

A. Output Range (Section 8.3)

OutPut

 

 

 

 

1.

4 mA Output

 

0 - 14 pH

 

0.00 pH

_______

2.

20 mA Output:

 

0 - 14 pH

 

14.00 pH

_______

3.

Hold

HOLd

3.80 to 22.00 mA

21.00mA

_______

4.

Dampening

dPn

0 to 255 seconds

0 seconds

_______

5.

Fault Current Output Setting

FAULt

3.80 to 22.00 mA

22.00mA

_______

B. Diagnostic (Section 8.4)

dIAgnOStIC

 

 

 

 

1.

Reference Cell Offset (Standardize error)

rOFFSt

0 to 1000 mV

 

60 mV (pH on

_______

 

 

 

 

 

glass electrode)

 

2.

Diagnostics Function

dIAg

On/Off

 

Off

_______

3.

Glass Impedance Temperature Correction

IMPtC

On/Off

 

On

_______

4.

Glass Electrode High Impedance Fault

GFH

0 to 2000 megohms

1500 megohms

_______

5.

Glass Electrode High Impedance Warning

GWH

0 to 2000 megohms

1000 megohms

_______

6.

Glass Electrode Low Impedance Warning

GWL

0 to 900 megohms

20 megohms

_______

7.

Glass Electrode Low Impedance Fault

GFL

0 to 900 megohms

10 megohms

_______

8.

Glass Impedance Calibration Warning

CAL

0 to 500 %

 

0 % (not Active)

_______

9.

Reference Cell Impedance Type

rEF

LO/HI

 

LO

_______

10. Reference Cell High Impedance Fault

rFH

0 to 2000 megohms (HI)

1500 megohms

_______

 

 

 

0 to 2000 kilohms

(LO)

140 kilohms

 

11. Reference Cell High Impedance Warning

rWH

0 to 2000 megohms (HI)

1000 megohms

_______

 

 

 

0 to 2000 kilohms

(LO)

40 kilohms

 

12. Reference Cell Low Impedance Warning

rWL

0 to 900 megohms

(HI)

20 megohms

_______

 

 

 

Does not apply for low impedance reference cell

 

13. Reference Cell Low Impedance Fault

rFL

0 to 900 megohms

(HI)

10 megohms

_______

 

 

 

Does not apply for low impedance reference cell

 

C. Temperature (Section 8.5)

tEMP

 

 

 

 

1. Auto Temperature Compensation

tAUtO

On/Off

 

On

_______

2.

Manual Temperature

tMAn

-15 to 130 0C

 

25 0C

_______

 

 

 

5 to 266 0F

 

 

 

3.

Temperature Sensor Type

tC

100-3; 100-4; 1000-3;

100-3

_______

 

 

 

1000-4: 3000

 

 

 

D. Display (Section 8.6)

dISPLAY

 

 

 

 

1.

Measurement type

tYPE

pH/ORP

 

pH

_______

2.

Temperature Units

tEMP

°C/°F

 

°C

_______

3.

Output Units

OUtPUt

mA/% of full scale

mA

_______

4.

Code

COdE

0 to 999

 

000

_______

E. Buffer (Section 8.7)

bUFFEr

 

 

 

 

1. Auto Calibration Function

b AUtO

ON/OFF

 

ON

_______

2.

Buffers Selection List

bUFFEr

See Tables 8-2 and 8-3

Standard

_______

3.

Auto Buffer Stabilization Time

tIME

0 to 30 seconds

 

10 seconds

_______

4. Auto Stabilization pH Change

PH

.002 to .5pH

 

.02 pH

_______

F. Isopotential (Section 8.8)

ISOPOtntAL

 

 

 

 

1.

Temperature Coefficient

tCOEF

-0.044 to 0.028 pH/ 0C

0.000 pH/ 0C

_______

2.

Solution Isopotential pH

ISO

-1.35 to 20.12 pH

7.00 pH

_______

3.

Sensor Isopotential pH

Snr

0.00 to 14.00 pH

7.00 pH

_______

G. Output Simulation (Section 8.9)

SIMOUtPUt

 

 

 

 

Test

tESt

3.80 to 22 mA

 

12.00 mA

_______

 

 

 

 

 

 

 