MODEL 3081 pH/ORP | SECTION 8.0 |
| PROGRAMMING FOR pH MEASUREMENTS |
TABLE 8-1. pH Settings List
ITEM | MNEMONIC | DISPLAY LIMITS | FACTORY SETTINGS | USER SETTINGS | ||
PROGRAM LEVEL (Sections 8.0 - 8.9). |
|
|
|
|
| |
A. Output Range (Section 8.3) | OutPut |
|
|
|
| |
1. | 4 mA Output |
| 0 - 14 pH |
| 0.00 pH | _______ |
2. | 20 mA Output: |
| 0 - 14 pH |
| 14.00 pH | _______ |
3. | Hold | HOLd | 3.80 to 22.00 mA | 21.00mA | _______ | |
4. | Dampening | dPn | 0 to 255 seconds | 0 seconds | _______ | |
5. | Fault Current Output Setting | FAULt | 3.80 to 22.00 mA | 22.00mA | _______ | |
B. Diagnostic (Section 8.4) | dIAgnOStIC |
|
|
|
| |
1. | Reference Cell Offset (Standardize error) | rOFFSt | 0 to 1000 mV |
| 60 mV (pH on | _______ |
|
|
|
|
| glass electrode) |
|
2. | Diagnostics Function | dIAg | On/Off |
| Off | _______ |
3. | Glass Impedance Temperature Correction | IMPtC | On/Off |
| On | _______ |
4. | Glass Electrode High Impedance Fault | GFH | 0 to 2000 megohms | 1500 megohms | _______ | |
5. | Glass Electrode High Impedance Warning | GWH | 0 to 2000 megohms | 1000 megohms | _______ | |
6. | Glass Electrode Low Impedance Warning | GWL | 0 to 900 megohms | 20 megohms | _______ | |
7. | Glass Electrode Low Impedance Fault | GFL | 0 to 900 megohms | 10 megohms | _______ | |
8. | Glass Impedance Calibration Warning | CAL | 0 to 500 % |
| 0 % (not Active) | _______ |
9. | Reference Cell Impedance Type | rEF | LO/HI |
| LO | _______ |
10. Reference Cell High Impedance Fault | rFH | 0 to 2000 megohms (HI) | 1500 megohms | _______ | ||
|
|
| 0 to 2000 kilohms | (LO) | 140 kilohms |
|
11. Reference Cell High Impedance Warning | rWH | 0 to 2000 megohms (HI) | 1000 megohms | _______ | ||
|
|
| 0 to 2000 kilohms | (LO) | 40 kilohms |
|
12. Reference Cell Low Impedance Warning | rWL | 0 to 900 megohms | (HI) | 20 megohms | _______ | |
|
|
| Does not apply for low impedance reference cell |
| ||
13. Reference Cell Low Impedance Fault | rFL | 0 to 900 megohms | (HI) | 10 megohms | _______ | |
|
|
| Does not apply for low impedance reference cell |
| ||
C. Temperature (Section 8.5) | tEMP |
|
|
|
| |
1. Auto Temperature Compensation | tAUtO | On/Off |
| On | _______ | |
2. | Manual Temperature | tMAn |
| 25 0C | _______ | |
|
|
| 5 to 266 0F |
|
|
|
3. | Temperature Sensor Type | tC | _______ | |||
|
|
|
|
|
| |
D. Display (Section 8.6) | dISPLAY |
|
|
|
| |
1. | Measurement type | tYPE | pH/ORP |
| pH | _______ |
2. | Temperature Units | tEMP | °C/°F |
| °C | _______ |
3. | Output Units | OUtPUt | mA/% of full scale | mA | _______ | |
4. | Code | COdE | 0 to 999 |
| 000 | _______ |
E. Buffer (Section 8.7) | bUFFEr |
|
|
|
| |
1. Auto Calibration Function | b AUtO | ON/OFF |
| ON | _______ | |
2. | Buffers Selection List | bUFFEr | See Tables | Standard | _______ | |
3. | Auto Buffer Stabilization Time | tIME | 0 to 30 seconds |
| 10 seconds | _______ |
4. Auto Stabilization pH Change | PH | .002 to .5pH |
| .02 pH | _______ | |
F. Isopotential (Section 8.8) | ISOPOtntAL |
|
|
|
| |
1. | Temperature Coefficient | tCOEF | 0.000 pH/ 0C | _______ | ||
2. | Solution Isopotential pH | ISO | 7.00 pH | _______ | ||
3. | Sensor Isopotential pH | Snr | 0.00 to 14.00 pH | 7.00 pH | _______ | |
G. Output Simulation (Section 8.9) | SIMOUtPUt |
|
|
|
| |
Test | tESt | 3.80 to 22 mA |
| 12.00 mA | _______ | |
|
|
|
|
|
|
|
57