Emerson Process Management 3081 pH/ORP instruction manual Orienting the Display Board

Models: 3081 pH/ORP

1 150
Download 150 pages 18.28 Kb
Page 24
Image 24

MODEL 3081 pH/ORP

SECTION 2.0

 

INSTALLATION

6.There are more jumpers on the CPU board. Refer to Figure 2-3. These jumpers are factory set and should NOT need to be moved. This step is for trou- bleshooting purposes only.

Verify that jumpers JP-1, JP-3, and JP-4 on the CPU board are in the positions shown in Figure 2-3. For installations where 50 Hz ac power is present, closing JP-3 may improve immunity of the trans- mitter to noise.

7.To reassemble the stack, place the dis- play board on the CPU board. Be sure the display board is properly oriented. The small window (the infrared detector for the remote controller) marks the top of the board. Insert the three bolts through the holes. Align the bolts with the stand- offs on the analog board and position the display and CPU boards on the analog board. If the boards are properly aligned, the bolts will drop in place. Press along the bottom of the stack to seat the pin and socket connector. Tighten the bolts.

8.Replace the end cap and lock nut.

2.3 ORIENTING THE DISPLAY BOARD

DWG. NO.

REV.

40008125

A

 

 

FIGURE 2-3. Model 3081 pH/ORP Transmitter CPU Board

The display board can be rotated 90 degrees, clockwise or counterclockwise, from the original position. To reposition the display:

1.Loosen the cover lock nut until the tab disengages from the circuit end cap. Unscrew the cap.

2.Remove the three bolts holding the circuit board stack.

3.Lift and rotate the display board 90 degrees, clockwise or counterclockwise, into the desired position.

4.Position the display board on the stand offs. Replace and tighten the bolts.

5.Replace the circuit end cap.

2.4MECHANICAL INSTALLATION 2.4.1 General information

1.The transmitter tolerates harsh environments. For best results, install the transmitter in an area where temperature extremes, vibrations, and electromagnetic and radio frequency interference are minimized or absent.

2.To prevent unintentional exposure of the transmitter circuitry to the plant environment, keep the security lock in place over the circuit end cap. To remove the circuit end cap, loosen the lock nut until the tab disengages from the end cap, then unscrew the cover.

3.The transmitter has two 3/4-inch conduit openings, one on each side of the housing. Run sensor cable through the left side opening (as viewed from the wiring terminal end of the transmitter) and run power/current loop wiring through the right side opening.

10

Page 24
Image 24
Emerson Process Management 3081 pH/ORP Orienting the Display Board, Mechanical Installation 2.4.1 General information