MODEL 3081 pH/ORP | SECTION 10.0 |
| PROGRAMMING FOR ORP MEASUREMENTS |
PROGRAM
rWJH | 0 | 040 |
EXIT |
| ENTER |
PROGRAM
rWJL | 0 | 000 |
EXIT |
| ENTER |
8. The rWH prompt appears. In the display, W appears as w j. Use the editing keys to change the display to the desired reference electrode high impedance warn- ing value. The allowed ranges are
Type of reference electrode | Allowed range | ||
|
|
| |
Low impedance (LO in step 6) | 0 | - 2000 kilohms |
|
|
|
| |
High impedance (HI in step 6) | 0 | - 2000 megohms |
Entering 0000 disables the feature. When the reference electrode impedance goes above the fault value, the transmitter displays the diagnostic message rEFWArn. Press ENTER to save.
9. The rWL prompt appears. Use the editing keys to change the display to the desired reference electrode low impedance warning value. The allowed ranges are
Type of reference electrode | Allowed range | |
|
|
|
Low impedance (LO in step 6) | not applicable | |
|
| |
High impedance (HI in step 6) | 0 - 900 megohms |
PROGRAM
rFL
EXIT
0000
ENTER
Entering 0000 disables the feature. When the reference electrode impedance goes below the warning value, the transmitter displays the diagnostic message rEFWArn. Press ENTER to save. The prompt appears but is disabled when LO is selected in step 6.
10.The rFL prompt appears. Use the editing keys to change the display to the desired reference electrode low impedance fault value. The allowed ranges are
Type of reference electrode | Allowed range | |
|
| |
Low impedance (LO in step 6) | not applicable |
|
|
| |
High impedance (HI in step 6) | 0 - 900 megohms |
Entering 0000 disables the feature. When the reference electrode impedance goes below the fault value, the transmitter displays the diagnostic message rEFFAIL and sets a fault condition. Press ENTER to save. The prompt appears but is disabled when LO is selected in step 6.
11. Press EXIT to return to the process display.