MODEL 3081 pH/ORP

SECTION 12.0

 

TROUBLESHOOTING

pH measurement problem

Perform buffer calibration (outside of the process)

Refer to Section 7.5

Buffer calibration

YES

Likely Process Problem

OK?

 

or Ground Loop:

NO

Place transmitter internal pream-

plifier switch in proper position.

NO

Is internal preamplifier switch in transmitter in correct position? See Figure 2-2.

1.Measure pH of a grab sample in a beaker. If pH is correct, then the process is OK.

2.Resolve the ground loop; insu- late shield wires from grounded metal. See Section 12.5.8

3.Ground un-grounded processes: (example: all plastic piping)

YES

Simulate pH directly into the trans-

mitter. Refer to Section 12.8.2.

 

 

 

 

 

 

 

 

 

Transmitter problem.

 

NO

Analyzer test OK?

YES

Likely Sensor Problem:

Replace Analog/CPU board

 

 

 

 

 

 

 

 

 

stack (PN 23574-02).

 

 

 

 

 

 

1.

Verify sensor is compatible with

 

 

 

 

 

 

 

 

the analyzer. Must have PT100,

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PT1000, or 3K Balco RTD. Verify

 

 

 

 

 

 

 

 

RTD jumpers and temperature

 

 

 

 

 

 

 

 

parameter programming.

 

 

 

 

 

 

 

2.

Check RTD. If bad or wrong, use

 

 

 

 

 

 

 

 

manual temperature compensa-

 

 

 

 

 

 

 

 

tion (see Section 8.5).

 

 

 

 

 

 

 

2.

Clean sensor & re-test it.

 

 

 

 

 

 

 

3.

If problem persists, rebuild or

 

 

 

 

 

 

 

 

replace the sensor.

FIGURE 12-6. Troubleshooting Flow Chart / Preamplifier in Transmitter or Built into Sensor

109

Page 123
Image 123
Emerson Process Management 3081 pH/ORP instruction manual Ok?