MODEL 3081 pH/ORP
PROGRAM
dIAGnOSTIC
EXIT NEXT ENTER
PROGRAM
rOFFSt | 0 | 60 |
|
EXIT |
| ENTER | |
|
|
|
|
|
|
|
|
PROGRAM |
|
|
|
d I A G |
|
| |
OFF |
| ||
EXIT |
| ENTER |
SECTION 10.0
PROGRAMMING FOR ORP MEASUREMENTS
10.4.3 Procedure
1. Press PROG on the infrared remote controller (IRC).
2. Press NEXT until the diAGnOStIC
3. The screen displays the rOFFSt prompt. Use the editing keys to change the flashing display to the desired standardization (reference) offset (in millivolts). The range is 0 to 1000 mV. Press ENTER to save.
4. The dIAG prompt appears. Use the Ï or Ð keys to enable (On) or disable (OFF) the sensor diagnostics. Press ENTER to save.
PROGRAM |
|
I M P t C | ON |
EXIT NEXT ENTER
PROGRAM
rEF | LO |
|
EXIT | ENTER |
5. The IMPtC prompt appears. Use the Ï or Ð keys to enable (On) or disable (OFF) glass impedance temperature compensation. Because glass impedance is a strong function of temperature, correcting glass impedance for temperature is recommended. A third set- ting (SPC) appears in addition to On and OFF. Do not select SPC; the setting is intend- ed for factory use. Press ENTER to save.
6. The rEF prompt appears. Press Ï or Ð until the desired setting appears. LO identifies a low impedance reference electrode, and HI identifies a high impedance reference electrode. Press ENTER to save. Selecting LO disables the low impedance warning and failure limits for the reference electrode.
NOTE
PROGRAM
rFH
EXIT
1400
ENTER
Be sure the jumpers on the analog board are set to match the reference electrode impedance. See Section 2.2,
7.The rFH prompt appears. Use the editing keys to change the display to the desired ref- erence electrode high impedance fault value. The allowed ranges are
Type of reference electrode | Allowed range | |
Low impedance (LO in step 6) | 0 | - 2000 kilohms |
High impedance (HI in step 6) | 0 | - 2000 megohms |
Entering 0000 disables the feature. When the reference electrode impedance goes above the fault value, the transmitter displays the diagnostic message rEFFAIL and sets a fault condition. Press ENTER to save.
80