SECTION 8.0

PROGRAMMING FOR pH MEASUREMENTS

8.5.3 Procedure

1. Press PROG on the infrared remote controller (IRC).

2. Press NEXT until the tEMP sub-menu appears in the display. Press ENTER .

3. The screen displays the tAUTO prompt. Press Ï or Ð to enable (On) or disable (OFF) automatic temperature compensation. Press ENTER to save.

4. The tMAN prompt appears. Use the editing keys to change the temperature to the desired value. To enter a negative number, press Î or Í until no digit is flashing. Then press Ï or Ð to display the negative sign. Permitted values are between -5.0 and 130.0°C. If tAUTO was disabled in step 3, the temperature entered in this step will be used in all subsequent measurements, no matter what the process tempera- ture is. Press ENTER to save.

PROGRAM

tC 100-3

EXITENTER

5.The screen shows the tC prompt. Press Ï or Ð to scroll to the desired temperature element and wiring configuration. Press ENTER to save.

1000-3

3 wire 1000 ohm RTD

1000-4

4 wire 1000 ohm RTD

100-3

3 wire 100 ohm RTD

100-4

4 wire 100 ohm RTD

3000

3000 ohm Balco RTD

NOTE

A jumper on the analog board must also be set to match RTD. See Section 2.2, Pre-Installation Set Up.

6. Press EXIT to return to the process display.