MODEL 3081 pH/ORP
1.3SPECIFICATIONS -GENERAL FOR MODEL 3081 pH/ORP
Case: Cast aluminum containing less than 8% magnesium. NEMA 4X (IP65), NEMA 7
Dimensions: 6.3 in. x 6.9 in. x 6.4 in. (160 mm x 175 mm x 161 mm); diameter 6.1 in (155 mm)
Conduit Openings: 3/4 in. FNPT
Reference Impedance: Transmitter accepts high impedance (i.e. glass) reference electrodes as well as low impedance (i.e.
Output:
Output can be programmed to go to any value between 3.8 and 22.0 mA to indicate a fault or hold condition.
Response Time: Display reaches 95% of final reading within 10 seconds.
Temperature Sensors: The following RTDs can be used with the Model 3081 pH/ORP transmitter:
3 and 4 wire Pt 100 RTDs
3 and 4 wire Pt 1000 RTDs
3000 ohm Balco RTD
Transmitter can also be used with
Local Display: Two line LCD; first line shows process vari- able (pH or ORP), second line shows temperature and output signal. When triggered, fault and warning mes- sages alternate with temperature and output readings.
Process variable: 7 segment LCD, 0.8 in. (20 mm) high
Temperature/output: 7 segment LCD, 0.3 in. (7 mm) high
Display board can be rotated 90 degrees clockwise or counterclockwise.
During calibration and programming, messages and prompts appear in the temperature/output area.
Power Supply and Load Requirements: See graph below. A minimum loop resistance of 250 Ω and a minimum power supply voltage of 18 Vdc is required for HART communication. Maximum power supply voltage for intrinsically safe and
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
Security: User selected security code prevents accidental changes to program settings.
Ambient Temperature:
Relative Humidity: 0 to 95% (with covers sealed)
Storage Temperature:
EMI/RFI: Meets the requirements of
Hazardous Area Classification:
Explosion Proof:
FM: | Class I, Div. 1, Groups B, C & D |
| Class II, Div. 1, Groups E, F, & G |
| Class III, Div. 1 |
CSA: | Class I, Div. 1, Groups C& D |
| Class I, Div. 2, Groups A, B, C & D |
| Class II, Div. 2, Groups E, F & G |
| Class III, Div. 1 |
Intrinsic Safety: | |
FM: | Class I, II & III, Div. 1 |
| T4 T AMB= 40°C; T3AT AMB= 70°C |
CSA: | Class I, Div. 1 |
T 3C T AMB=40°C; T3 T AMB=80°C CENELEC: EEx ia IIC
T5 Tamb=40°C; T4 Tamb=65°C
Non-Incendive:
FM: Class I, Div. 2, Groups A, B, C & D
CSA: Class I, Div. 2, Groups A, B, C & D T5 (Tamb=40°C)
Weight/Shipping Weight: 10 lb/10 lb (4.5 kg/4.5 kg).
Weights and shipping weights are rounded to the nearest whole pound.
minimum voltage and load for HART communication