MODEL 3081 pH/ORP | SECTION 8.0 |
| PROGRAMMING FOR pH MEASUREMENTS |
SECTION 8.0
PROGRAMMING FOR pH MEASUREMENTS
8.1General
8.2Entering and Leaving the Program Menu
8.3Output Ranging
8.4Diagnostic Parameters
8.5Temperature Related Settings
8.6Display Units
8.7Buffer Calibration Parameters
8.8Isopotential Parameters
8.9Generating a Test Current
8.1 GENERAL
This section describes how to do the following:
1.assign pH values to the 4 and 20 mA outputs,
2.set the current generated by the transmitter during hold,
3.set the current generated by the transmitter when a fault is detected,
4.change sensor diagnostic limits,
5.enable and disable automatic temperature compensation,
6.change the units of the displayed variables,
7.program a security code,
8.identify buffers for auto calibration,
9.change the transmitter isopotential point,
10.simulate output currents for testing.
Factory default settings are given in Table
8.2 ENTERING AND LEAVING THE PROGRAM MENU
Press PROG on the infrared remote controller (IRC) to enter the Program menu. To save new settings, press ENTER. To leave the Program menu without saving new values, press EXIT . Pressing EXIT with a prompt showing returns the display to the first prompt in the
If program settings are protected with a security code, pressing PROG or CAL will cause the Id screen to appear. Key in the security code and press ENTER . The first
A transmitter adjacent to the one being programmed may pick up signals from the IRC. To avoid accidentally changing settings, use a different security code for each nearby transmitter. See Section 5.7, Security, and Section 8.6, Display Units, for details.
56