MODEL 3081 pH/ORP | SECTION 8.0 |
| PROGRAMMING FOR pH MEASUREMENTS |
8.7 BUFFER CALIBRATION PARAMETERS
8.7.1 Purpose
This section describes how to do the following:
1.activate or deactivate auto calibration,
2.identify which buffers will be used during auto calibration,
3.set the stability criteria for auto calibration.
8.7.2 Definitions
1.AUTO CALIBRATION. In auto calibration, screen prompts direct the user through a two point buffer calibration. The transmitter recognizes the buffers and uses
2.BUFFERS. Buffers are aqueous solutions to which exactly known pH values have been assigned. Assigning a pH value to a buffer involves certain fundamental assumptions. Slightly different assumptions lead to slightly different pH scales. Over the years, various national standards organizations have developed different scales. The Model 3081 pH/ORP transmitter recognizes the common standard scales as well as common commercial buffers. Commercial buffers, which are sometimes called technical buffers, are traceable to standard buffers, but the accuracy of commer- cial buffers is generally less than standard buffers. Tables
TABLE
|
| NIST |
| DIN 19266 |
| JIS 8802 |
| BSI |
| |||
| pH |
| temp (°C) | pH |
| temp (°C) | pH |
| temp (°C) | pH | temp (°C) |
|
| 1.68 |
| 5 - 95 | 1.68 |
| 5 - 95 | 1.68 |
| 0 - 95 | 1.68 | 0 - 60 |
|
| 3.56 |
| 25 - 95 |
|
|
|
|
|
| 3.56 | 25 - 60 |
|
| 3.78 |
| 0 - 95 |
|
|
|
|
|
|
|
|
|
| 4.01 |
| 0 | 4.01 |
| 0 - 95 | 4.01 |
| 0 - 95 | 4.01 | 0 - 60 |
|
| 6.86 |
| 0 | 6.86 |
| 0 - 95 | 6.86 |
| 0 - 95 | 6.86 | 0 - 60 |
|
| 7.00 |
|
|
|
|
| see note |
|
|
| ||
| 7.41 |
| 0 - 50 |
|
|
|
|
|
|
|
|
|
| 9.18 |
| 0 - 95 | 9.18 |
| 0 - 95 | 9.18 |
| 0 - 95 | 9.18 | 0 - 60 |
|
| 10.01 |
| 0 - 50 |
|
|
| 10.01 |
| 0 - 50 | 10.01 | 0 - 50 |
|
| 12.45 |
| 0 - 60 | 12.45 |
| 0 - 60 |
|
|
|
|
|
|
NOTE: pH 7 buffer is not a standard buffer. Because it is a popular commercial buffer in the United States, it is included with the standard buffers. The pH of the buffer is defined between 0 and 95°C.