MODEL 3081 pH/ORP

PROGRAM

OutPut

EXIT NEXT ENTER

PROGRAM

4MA

 

-

1 4 0 0

EXIT

NEXT

 

 

 

 

ENTER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PROGRAM

 

 

 

 

 

 

20MA

 

 

 

400

 

1

EXIT

NEXT

 

 

 

 

ENTER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PROGRAM

 

 

 

 

 

 

HoLd

 

 

1 . 0 0

 

2

EXIT

NEXT

 

 

 

 

ENTER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PROGRAM

 

 

 

 

 

 

FAULt

 

 

2.00

 

2

EXIT

NEXT

 

 

 

 

ENTER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PROGRAM

 

 

 

 

 

 

dPn

 

 

 

 

 

00

 

0

EXIT

NEXT

 

 

 

 

ENTER

 

 

 

 

 

 

 

 

SECTION 10.0

PROGRAMMING FOR ORP MEASUREMENTS

10.3.3 Procedure

1. Enter the Program menu by pressing PROG on the IRC. The OutPut sub-menu appears.

2. Press ENTER. The screen displays the 4 MA prompt. Use the editing keys to change the displayed number to the desired ORP. The allowed range is -1400 to 1400. To change the display to a negative number, press Î or Í until no digit is flashing. Then press Ï or Ð until the minus sign appears. To change the display to a positive number, press Î or Í until the negative sign is flashing. Then press

Ïor Ð until the minus sign disappears. Press ENTER to save.

3. The screen displays the 20 MA prompt. Use the editing keys to change the displayed number to the desired ORP. The allowed range is -1400 to 1400. Press ENTER to save.

4. The screen displays the HoLd prompt. Use the editing keys to change the display to the output desired when the transmitter is in hold. The range is 3.80 to 22.00 mA. Entering 00.00 causes the transmitter to hold the output at the value it was when placed in hold. The hold setting overrides the fault setting. Press ENTER to save.

5. The screen displays the FAULt prompt. Use the editing keys to change the display to the output desired when the transmitter detects a fault. The range is 3.80 to 22.00 mA. Entering 00.00 causes the transmitter to hold the output at the value it was when the fault occurred. Press ENTER to save.

6. The screen displays the dPn prompt. Use the editing keys to change the display to the desired output dampening. The range is 0 to 255. Press ENTER to save.

7. Press EXIT to return to the process display.

78

Page 92
Image 92
Emerson Process Management 3081 pH/ORP instruction manual OutPut, 20MA 400