WHAT YOU NEED TO KNOW

BEFORE INSTALLING AND WIRING A ROSEMOUNT ANALYTICAL

SENSOR TO THE MODEL 3081 pH/ORP TRANSMITTER

1.THE MODEL NUMBER OF THE SENSOR

Look on the label.

Also note the model option string.

If the label is missing or unreadable, see the flowcharts on pages 28 through 30.

Write the sensor model number here

2.THE TYPE OF TEMPERATURE ELEMENT

Look on the label.

If the label is missing or unreadable, measure the resistance between the RTD leads.

Write the temperature element RTD here

 

If resistance is . . .

 

the RTD is . . .

 

 

 

 

 

about 110 ohms

 

Pt 100

 

 

about 3000 ohms

 

Balco 3K

 

 

 

 

 

 

3.THE LOCATION OF THE PREAMPLIFIER: INSIDE OR OUTSIDE THE TRANSMITTER?

If the sensor is wired through a junction box, the preamplifier is ALWAYS in the junction box or the sensor.

Page 3
Image 3
Emerson Process Management 3081 pH/ORP instruction manual What YOU Need to Know